Medtronic identified that the Pelvic Health clinician apps, which are installed on the Smart Programmer mobile device, have a password vulnerability that requires a security update to fix. Not updating could potentially result in unauthorized control of the clinician therapy application, which has greater control over therapy parameters than the patient app. Changes still cannot be made outside of the established therapy parameters of the programmer. For unauthorized access to occur, an individual would need physical access to the Smart Programmer.
Detalles CVE
Puntuacion CVSS v3.16.4
SeveridadMEDIUM
Vector CVSSCVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:L/I:H/A:H
Vector de ataquePHYSICAL
ComplejidadLOW
Privilegios requeridosNONE
Interaccion usuarioNONE
Publicado3/1/2023
Ultima modificacion11/21/2024
Fuentenvd
Avistamientos honeypot0
Productos afectados
medtronic:interstim_x_clinicianmedtronic:micro_clinician
Debilidades (CWE)
CWE-620CWE-287
Referencias
https://global.medtronic.com/xg-en/product-security/security-bulletins/pelvic-health-interstim-micro.html(security@medtronic.com)
https://global.medtronic.com/xg-en/product-security/security-bulletins/pelvic-health-interstim-micro.html(af854a3a-2127-422b-91ae-364da2661108)